2 flagged images · first reported 24 July 2026 · by Reese Richardson
What was found
- Background matches background pattern C
- Background matches background pattern D (partial match to background pattern C)
Background pattern C, D — the identical noise field appears in thousands of otherwise unrelated blots, which is not possible for independently acquired images.
The images
Annotations by the contributors listed above, republished under CC BY 4.0 from the Zenodo repository.
Other products carrying the same background
- MA5-31646 Thermo Fisher Scientific — BCL11B Monoclonal Antibody (1F8G8)
- ab233710 Abcam —
- NBP2-37337 R&D Systems — TRACP/PAP/ACP5 Antibody (7E6A11) - BSA Free
- ab233720 Abcam —
- MA5-31691 Thermo Fisher Scientific — CHRNA7 Monoclonal Antibody (7F10G1)
- MA5-17172 Thermo Fisher Scientific — SFTPC Monoclonal Antibody (5E6A9)
A flag on this page means the images the vendor published as evidence cannot be trusted as evidence. Unless an independent knockout test is shown above, it says nothing about whether this antibody binds its target. Treat it as untested until you or someone else controls it.