Search / Thermo Fisher Scientific
MA5-15915
DLK1 Monoclonal Antibody (3A10)
1 flagged image · first reported 23 July 2026 · by Reese Richardson
What was found
- Background matches background pattern C
Background pattern C — the identical noise field appears in thousands of otherwise unrelated blots, which is not possible for independently acquired images.
The images
Annotations by the contributors listed above, republished under CC BY 4.0 from the Zenodo repository.
Other products carrying the same background
- MA5-31803 Thermo Fisher Scientific — Myostatin Monoclonal Antibody (6E4E6)
- MA5-15916 Thermo Fisher Scientific — UBC9 Monoclonal Antibody (1B10)
- MA5-17028 Thermo Fisher Scientific — CD166 Monoclonal Antibody (5B3)
- MA5-15927 Thermo Fisher Scientific — NQO1 Monoclonal Antibody (1A11)
- MA5-15876 Thermo Fisher Scientific — LIMS1 Monoclonal Antibody (5G7)
- MA5-17195 Thermo Fisher Scientific — TWIST1 Monoclonal Antibody (2F8E7)
A flag on this page means the images the vendor published as evidence cannot be trusted as evidence. Unless an independent knockout test is shown above, it says nothing about whether this antibody binds its target. Treat it as untested until you or someone else controls it.